7100P Portable Trace Oxygen Analyzer

Trace Oxygen Analyzer

The Neutronics Model 7100P Portable Trace Oxygen Analyzer is designed for portable trace oxygen gas measurement in pure gases such as nitrogen, argon and helium.

  • 0 to 1000 ppm measurement range
  • Mini-zirconia oxide ZR-400 sensor
  • Pump drive for extractive or positive pressure sampling
  • Sample flowmeter for easy-to-read monitoring
  • Simple start-up
Request More Information
Download Datasheet
Categories: , Tag:

Description

The Neutronics 7100P Portable Trace Oxygen Analyzer is a portable analyzer designed for trace oxygen gas measurement in pure gases such as nitrogen, argon, and helium. This analyzer features a rapid-response mini-zirconia sensor with a measurement range of 0 to 1000 ppm oxygen. Extremely fast response, high accuracy, and simple calibration make this portable analyzer a low-maintenance solution that delivers reliable performance in a rugged light-weight steel enclosure.

The ZR-400 sensor is a non-depleting solid state device based on a zirconia electrolyte cell. With an integral heater and two electrodes, it is mounted directly onto a printed circuit board that supplies bias and heater voltages to the sensor element. A small capillary on the sensor surface controls diffusion of oxygen into the sensor. At operating temperature, oxygen is electrochemically reduced, causing current flow through the solid electrolyte. Oxygen concentration in the sample gas is determined by measuring the current flowing through the electrodes.

  • Product Specifications
  • Technical Library
  • Part Numbers
  • Gas List
Measuring Range: 0-100 ppm, 0 - 1000 ppm auto
Accuracy: ±2.0% of range at CTP or ±3 ppm
Enclosure Rating: Ctrl panel: WeathrproofNEMA 4 (IP66)Rear electronic chassis:Nema 1, IP20
Response Time: T90 < 10 sec
Display Size: 7 seg. 0.75" LED4 alphanumeric characters;LEDs for system status: run,/fault/alarm 1/alarm 2
Power: 90 - 264 VAC or 24 VDC
Analog Communications: Serial RS-232 service portAnalog 0-1, 0-5,0-10 VDCand 4-20 mA
Relay Rating: 2 alarm relays, field adjustable 1 system fault relay, non-adj
Ambient Temperature: 41 to 104°F (5 to 40°C)
Humidity: 0 to 95% RH (non-condensing)
Sensor Type: Mini-Zirconia ZR400
Expected Sensor Service Life: 3 to 4 Years
Warranty: Analyzer: 1 YearSensor: 6 months
Description Part Number
7100P-CE, 90-264 VAC C7-01-7100-01-0
7100P, 24 VDC C7-01-7100-01-2